Dual Lumen Bolt Kit

//Dual Lumen Bolt Kit
Request A Quote
  • Provides dual port access
  • Other than a drill, requires no additional tools to insert or remove the bolt
  • Made of titanium
  • Secures the perfusion probe and additional sensor
  • Clear lumens to facilitate reading sensor depth markings
  • Wing nut for easy threading
  • CT Compatible
  • MR Conditional for static magnetic field of 3.0 T or less

The kit contains:

  • Dual Lumen Bolt
  • Compression Fittings
  • Sealing Caps
  • Stylet
  • Introducer
  • 5.3 mm Drill Bit with Depth Collar
  • Hex Wrench
  • Scalpel
  • Pigtail Extension
  • Male-Female Luer Coupler

What are the sizes of the lumens?

  • The long lumen can accommodate devices that are less than 0.089″ in diameter. The short Lumen can accommodate devices that are less than 0.053″ in diameter.

Is the Dual Lumen Bolt CT compatible?

  • Yes

 Dimensionally Compatible Sensors*

  • QFlow 500TM Perfusion Probe
  • Codman Microsensor ICP
  • Sophysa ICP (PSO-PTT)
  • Depth Electrode(s)
  • Raumedic Neurovent (PTO 2L)
  • Natus Camino ICP (1104-L)
  • Integra Licox T+P02 (CC1P1)
  • Integra Licox PbT02 (CC1SB)
  • 70 & 71 Microdialysis Bolt Catheter

*Hemedex makes no claims with respect to the functionality and performance of any sensor other than the Hemedex QFlow 500™ Perfusion Probe.

Can the Dual Lumen Bolt be used in an MR environment?

  • Non-clinical testing has demonstrated the Hemedex QFlow 500™ Titanium Bolt Kit is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:
    •  Static magnetic field of 3.0 T or less;
    • Maximum spatial field gradient of 8,610 G/cm (86.1 T/m);
    • Maximum MR system-reported, whole-body averaged specific absorption rate (SAR) of 2 W/kg (Normal Mode).
  • Under the scan conditions defined above, the Hemedex QFlow 500™ Titanium Bolt Kit is expected to produce a maximum temperature rise of less
    than 0.3 ºC after 15 minutes of continuous scanning at 1.5 T and less than 0.6 ºC after 15 minutes of continuous scanning at 3 T.
  • In non-clinical testing, the image artifact caused by the device extends approximately 25 mm from the Hemedex QFlow 500™ Titanium Bolt when
    imaged with a gradient echo pulse sequence and a 3 T MRI system.
  • Prior to transporting the patient to the MRI suite, remove all MR unsafe/conditional sensors that are not labelled for use in an MR environment.
    • NOTE: Never leave Stylets or MR unsafe/conditional sensors that are not labeled for an MR environment in the bolt during MRI.
  • Hemedex recommends that the lumens be closed/capped following the removal of any sensors.

Keep up with our latest news:

2020-07-13T17:36:26+00:00